International Journal of Research and Scientific Innovation (IJRSI)

Submission Deadline-23rd December 2024
Last Issue of 2024 : Publication Fee: 30$ USD Submit Now
Submission Deadline-05th January 2025
Special Issue on Economics, Management, Sociology, Communication, Psychology: Publication Fee: 30$ USD Submit Now
Submission Deadline-20th December 2024
Special Issue on Education, Public Health: Publication Fee: 30$ USD Submit Now

Real-Time Radiological Practice: Dose Descriptors and Fatal Risk of Cancer Estimates in Adult Patient from Most Frequent Radiological Procedures in Ondo State, Nigeria

Real-Time Radiological Practice: Dose Descriptors and Fatal Risk of Cancer Estimates in Adult Patient from Most Frequent Radiological Procedures in Ondo State, Nigeria

Fredrick O. Adeyemi1*, Christopher J. Olowookere2

1Department of Physical Sciences, Olusegun Agagu University of Science and Technology, Okitipupa, Ondo State, Nigeria

2Department of Physics, University of Medical Sciences, Ondo -Town, Ondo State, Nigeria

*Corresponding Author

DOI: https://doi.org/10.51244/IJRSI.2023.1012018

Received: 06 November 2023; Revised: 15 November 2023; Accepted: 20 November 2023; Published: 05 January 2024

ABSTRACT   

Background: The inspiration in establishing dosimetric methods for radio-diagnosis is based on the curiosity to estimating health risks to patient subjected to a given type of examination, using ionizing radiation. So, to establish strong relation between risks and benefits of a new diagnostic modality, study assessed real-time radiological practice by evaluating dose descriptors and fatal risks of cancer from most frequent radiographs.

Patients and Methods: Mathematical approach was adopted to determine dose to 3,587 adult patients while PCXMC version 2.0, was employed in estimating their effective dose equivalent.

Results: Low ESD and DAP recorded as 0.618 mGy (Ankle) and 0.124 Gycm2 (Hand AP) while high as 6.263 mGy (LS LAT) and 2.866Gycm2 (LS AP). The average estimated effective doses across radiographs confirmed Knee joint LAT and Ankle AP recording same low value of 0.001mSv and high as 0.822mSv from LS LAT.  The range factor for the probability of fatal risks of cancer evaluated across radiographs was 10.142. Thus, low average lifetime risks of cancer ratio recorded as 1:27,778(Ankle AP) and high for LS LAT as 1:2,739, and the risks of cancer inducement estimates from this study fall within very low and low category of risks. Comparison of study results with some published values show good geometry but revealed high spread in values across radiographs studied. The study trend demonstrates the significance for awareness creation among the radiographic staff and the policies maker on quality control testing of equipment and adjustment of protocols.

Conclusion: The need to bring the radiological practice in the State in line with National and European guidelines on quality criteria for diagnostic radiological images becomes very essential, for good radiological practice to be embraced.

Keywords: Real-time radiological practice, Dose descriptors, fatal risks of Cancer, Radiological procedures

INTRODUCTION

X-rays usage in modern medical practice is requisite, and has been clinically used for diagnosis and therapeutic purposes. This, continuously play a leading and significant role in spite of other imaging techniques both in developed and developing areas, globally. [1][2]. However, x-ray examination had constituted the most important man made source of radiation exposure of the world population [3]. Although, x-ray procedure are design to provide net benefit, the potential for radiation induced injury to the patient, staffs and the environment, and the deleterious effect increased with dose. The contribution of diagnostic radiology towards the annual radiation dose to population has been well documented [4]. So, understanding the absorbed dose and factors that affect these, becomes very noteworthy [5].

In view of the important benefit to patients from properly conducted medical exposures, the radiological protection concern focused on reduction of the irrational exposure, which are examination that are either unlikely to be helpful to the patient management or involve doses that are not ‘As Low As Reasonably Achievable’ (ALARA) in order to meet specific clinical objectives. So, reduction of unnecessary exposure will transform to reduction of deleterious effects associated with the use of the atom, and this embrace the justification for optimization process in patient dosimetry.

In diagnostic radiology, assessment of patient dose and quality control test on x-ray machine are required periodically to ensure compliance with recommendations [6][7][8]. So, the major lesson learned from these valuations is the recognition of the significant variations in patient’s dose among radiological department for same type of examinations [9], and these variations, be seen as justifying dose measurement in order to optimize diagnostic practice [10]. Thus, the need for radiological practice and dose to patient assessment arises, especially among the identifiable radiation least monitored areas, as specified by the Nigerian Nuclear Regulation Authority (NNRA). So, this study assesses radiation dose descriptors and fatal risks of cancer from frequently performed x-ray examinations and compare its average estimated parameters values, as a standard to patient from real-time radiological practice in the state with some local and internationally published values. Comparison of obtainable values becomes relevant, since there is no acceptable local diagnostic reference limits established in the state, as the time of study for evaluation.

MATERIALS AND METHODS

2.1.   Study area

Ondo state is a developing state within the Southwestern region of a developing country, Nigeria. It lies amid longitudes 40 30” and 6” east of the Greenwich meridian, 50 45” and 80 15” north of the equator, meaning that it entirely in the tropics. It has a population of about 3.46 million, comprising of 1.75 and 1.72 million for male and female respectively as at 2006 (Ondo State bureau of statistics) with land mass / area of about 14,788.732 km2 [11]. It made up of 18 local government areas and healthcare establishment, comprises of Federal medical centre, State specialist/general hospitals, Comprehensive health centres, and

Registered private hospitals which sum up to 223 in numbers. Present in each local government of the state is at least one specialist/general hospital, comprehensive, and basic healthcare centres. In the state, no assurance that patient protection during routine x-ray examinations has been given much attention nor any local diagnostics reference limits, established.

2.2. Study sample and sampling technique

This study was conducted in seventeen (17) selected hospitals, spread across eleven (11) of the 18 local governments in the State, with appropriate Health Research and Ethics Committee approval issued from the State Ministry of Health (Reg. no. ODMOHREC 06/07/2021). The selected hospital includes eleven (11) public (i.e. State Specialist/General Hospital (SSH)-7, Federal Medical Centre (FMC) – 2, Teaching Hospital (GFHDC)-1, a higher institution of learning Health Centre (OAUSTECH)-1) and six (6) registered private hospitals coded as indicated in Table 1, with criteria for choice of diagnostic centre includes good representation of x-ray procedure studied, geographical location, and hospital workload which are recommendable [12]. Centre’s x-ray machine information was considered appropriately (Table 1) and the exposure parameters used during clinical examinations were duly recorded. Statistical population of 3,587 adult patients exposed for different radiographs [13], considered and respective patient’s characteristics information recorded. The basic methodology adopted for this study was indirect dosimetry approach, supported by IAEA and ICRP in the absent of the use of thermoluminiscense dosemeter (TLD). So, quantities required for ESD and DAP collated for clinically examined patient including their exposure factors as tube loading (mAs), tube potential (kVp), focus to skin distance (FSD), focus to film distance (FFD), beam size (field size area) together with x-ray machine beam output Do ( ). The beam output of each machine were determined using x-ray test device (Noninvasive evaluation of radiation output, model 4000TM Victoreen Inc. USA), and as at the time of this study, the calibration of the x-ray test device as issued by the manufacturer was still valid. The reproducibility and linearity of the x-ray machines were checked at focus to detector distance of 1metre with constant tube voltage potential of 80 kVp (voltage at which anode current is assumed to be stabled) and tube loading of 10 mAs. The measured beam output in  was converted to  using conversion factor of 8.73×10-3 [13], and values (beam output) was used alongside others in equations 1 and 2 for patient ESD and DAP estimates. So, this study determined radiation dose of patient that has undergone radiographic examinations and judged locally at each selected centre by individual radiographer to have achieved the required image quality for diagnosis. Here, the performance of the manufacturer’s cassettes with screen-film combination speed ranged (between 200 and 800) was not assessed, since study aimed at estimating patient dose based on their anatomical data and exposure parameters used during clinical examination.

2.3.   Patient dose assessment

Study’s patient entrance skin surface dose (ESD) was calculated using [14] [15]:

ESD = Do (mGy/mAs).q (mAs). (FDD/FSD). (kV/80)2. f -1                                                                      1

Where, f, the backscattered factor used (‘f’ for standard adult =1.35), ‘kV’ and ‘q’ are the tube potential voltages and tube loading respectively, used during clinical examination. The respective dose-area product (DAP) was determined using [16]:

DAP = Do (mGy/mAs). q (mAs) . (A)FSD(cm)²                                                                                      2

Where, (A)FSD is the cross  sectional area of the x-ray beam on patient skin, which was determined by [16]:

(A)FSD =(FSD/FFD)2 A (FFD)                                                                                                            3

Where, A (FFD) is the field size area. However, equation 3 holds, when inverse square law is not considered during quality control measurement and consequently, measured DAP on patient’s skin. Accredited dosimetric software (PCXMC version 2.0), for calculating patient dose in medical x-ray examinations, designed by STUK (Radiation and Nuclear Safety Authority, Helsinki, Finland) was employed to estimates patient’s effective doses (ED). The software generate x-ray spectrum, based on the real input parameters (kVp, anode angle setting, total filtration, etc). During stimulation, maximum energy setting was 150 keV for the required number of photons. The dose calculation method by the software was based on Monte Carlo stimulation, which was in accordance with the stochastic model of interactions between photons and matter. So, PCXMC calculated effective dose (ED) to patient, using anatomical data from the mathematical phantom models, based on tissues weighting factors of ICRP publication 103[17]. The required input data for these calculation includes: definition of all projections (location and size of the radiation field and projection angle) and the exposure factors related to different x-ray examinations. The performance and stimulation of each x-ray projection was based on standard guidelines. On inputting the required data, resultant effective doses (ED) at the radiation field were obtained for every projection for each examination, subject to anode angle setting of 100 and average filter length of 2.5mmAl which was made constant. Selected skin and focus points was made constant for each radiograph by sex. Subsequently, calculated ESD was converted to tissue dose equivalent (HT) using conversion factor of 1.06 [8], which was used to assessed probability of fatal cancer incidence for each radiograph by sex and averaged over sexes, using [18]:

                                                                                                           4

Where, Rf is the probability of fatal risk of cancer incidence and   the risk of fatal cancer inducement for adult population (5.5×10-5 mSv-1). Patient related data were collated during clinical examination, to provide real-time feedback on relative magnitudes to the radiographer during process and influence the conduct of the procedure, so as to collate the quantity of the most interest (AAPM, 2002) [19]. So, result presented in this study was made averaged over groups by sex and averaged over sexes for the representation of the state result at 95% CL, suitable for risk analysis and diagnostic dose reference limit creation [17].

RESULTS

Table 1(a): X-ray machines specific radiographic data for Group A selected centres studied

Hospitals X-ray tube model Year of Installation Filter length (mmAl) Beam Output (µGy/mAs) Film type Film screen speed Servicing frequency
SSHK Ralco (Static) 2005 2.5 44.71 Kodak 200 On intervention of repair
SSHR Ralco (Static) 2005 2.5 46.55 Kodak 200 On intervention of repair
SSHO Ralco (Static) 2005 2.5 40.10 Kodak 200 On intervention of repair
SSHA Ralco (Static) 2005 2.5 49.05 Kodak 200 On intervention of repair
SSHID Ralco (Static) 2009 2.5 39.82 Kodak 200 On intervention of repair
SSHON Ralco (Static) 2005 2.5 51.42 Kodak 200 On intervention of repair
SSHIGB Ralco(static) 2010 2.5 48.22 Kodak 200 On intervention of repair
     FMCO Ralco(mobile) 2013 2.0 38.59 Kodak 400 Yearly
FMCA Ralco(static) 2013 2.5 43.11 Agfa 200 Yearly
OAUSTECH GEC Medical(static) 2014   1.5*  74.28** Kodak 400 On intervention of repair
GFHDC Roentgen 501 (siemen) (static) 2009 2.5 25.05 Carest Ream 800 Bi-annually

Table 1(b): X-ray machines specific radiographic data for Group B selected centres studied

Hospital X-ray tube model Year of installation Filter length  (mmAl) Beam output (µGy/mAs) Film type Film screen speed Servicing frequency
HSMCA Rountge 501 (125kv) static mAs 2013 2.5 35.78 Agfa 400 Yearly
BNMRSO Reta practice (philips) mobile (125kV, 500mAs) 2011 2.0 51.10 Kodak 200 Yearly
NHMO GEC (MX-4) static 125kV,500mAs 2004   1.7* 45.35 Kodak 200 Yearly
KMCA Philip Capasitor Discharge mobile (100kV) 2012   1.5* 45.91 Agfa 400 Bi-annually
MEDRO Mediroll-4, Neodic fnomax (medicor) static 125 kV , 500mAs 2015 2.5 35.47 Carest ream 800 Quarterly
TCHO Roentge (sienens) static 125kV, 500mAs 2013 2.5 37.44 Agfa 400 Bi-annually

Note: All centre x-ray machine studied were single phase with either 1 or 2 pulse/s, and used grid (r = 12/ 40 ).

*X-ray machine with low filter length;

**X-ray machine with High beam output (Recommended beam output for a single phase x-ray   machine generally ranged as [(4.0 ± 1.5)   = (35.92 ± 13.10) (µGy)/mAs)] [20] [21].

Table 2: Patients’ characteristics information by sex, averaged over sexes and radiographs, presented in mean (male and female), and averaged over sexes in mean and SEM                                           (in bracket)

Radiograph                    Sex Age (yrs.) Weight (WT) (kg) Height (Ht) (cm)  BMI (kgm-2) Depth (tcm)
Male Female Both  Male Female Both Male Female Both  Male Female Both Male Female Both Male Female Both
Chest PA 221 349 570 40.5

(7.5)

35.5

(11.5)

38.0

(2.5)

63.3

(5.1)

56.4

(4.3)

59.9

(3.5)

159.9

(7.5)

154.9

(7.9)

157.4

(2.5)

24.8

(0.4)

23.5

(0.6)

24.2

(0.7)

22.5

(0.4)

21.5

(0.3)

22.0

(0.5)

Chest AP 33 43 76 28.3

(4.1)

42.6

(8.5)

35.0

(7.2)

58.1

(0.0)

50.6

(0.0)

54.4

(3.8)

161.7

(0.0)

158.3

(0.0)

160.0

(1.1)

22.1

(1.9)

20.2

(2.0)

21.2

(1.0)

21.4

(1.0)

20.2

(1.0)

20.8 (0.6)
ABD AP 83 140 223 31.6

(3.6)

30.3

(2.8)

31.0

(0.7)

63.9

(0.9)

64.2

(7.0)

64.

(0.2)

159.6

(8.5)

158.8

(0.6)

159.2

(0.4)

25.3

(2.4)

25.5

(2.6)

25.4 (0.1) 22.6 (0.5) 22.7

(1.2)

22.7 (0.1)
Pelvis AP 38 80 118 46.5

(0.0)

36.5

(1.6)

41.5

(5.0)

63.7

(2.5)

60.8

(17.7)

62.3

(1.5)

160.1

(8.2)

155.2

(6.0)

157.7

(2.5)

24.9

(1.6)

24.8

(5.4)

24.9

(0.1)

22.6

(0.2)

22.1

(2.9)

22.4 (0.3)
Pelvis LAT 38 80 118 46.5

(0.0)

36.5

(1.6)

41.5

(5.0)

63.7

(2.5)

59.5

(16.4)

61.6

(2.1)

160.1

(8.2)

156.2

(7.0)

158.2

(2.0)

24.9

(1.6)

24.8

(5.4)

24.9 (0.1) 34.5

(0.8)

31.7

(2.6)

33.1 (1.4)
Skull (AP/PA) 40 26 66 24.8

(3.5)

41.4

(1.3)

33.1

(8.3)

51.4

(0.4)

56.7

(7.0)

54.1

(2.7)

154.2

(6.0)

150.4

(3.2)

152.3

(1.9)

21.7

(1.5)

24.0

(2.1)

23.4 (1.7) 18.8 (0.3) 17.9

(1.2)

18.4 (0.5)
Skull LAT 11 3 14 25.2

(3.2)

40.1

(4.8)

32.7

(7.5)

56.8

(5.8)

63.1

(0.6)

60.0

(3.2)

155.7 (7.5) 153.5

(0.0)

154.6

(1.1)

23.4

(0.2)

27.0

(0.1)

25.2 (1.8) 16.1 (9.3) 11.2

(7.2)

13.7 (2.5)
Hand AP 83 124 207 21.2

(0.9)

33.8

(7.2)

27.5

(6.3)

62.6

(5.4)

54.1

(6.1)

58.4

(4.3)

160.2

(7.0)

146.0

(3.5)

153.1

(7.1)

24.7

(1.8)

25.6

(4.1)

25.2 (0.5) 8.1 (0.2) 6.7

(1.5)

  7.4

0.7)

Knee Joint (LAT) 146 139 285 43.1

(4.1)

29.6

(8.6)

36.4

(6.8)

58.6

(3.6)

55.1

(0.1)

56.9

(1.8)

162.3

(7.1)

158.7

(11.4)

160.5

(1.8)

24.8

(2.5)

22.2

(3.2)

23.5 (1.3) 14.7

(0.4)

11.3

(1.2)

13.0 (1.7)
LSJ AP 102 191 293 47.5

(0.5)

53.1

(1.1)

50.3

(2.8)

71.2

(0.9)

65.3

(3.0)

68.3

(3.0)

162.1

(1.0)

161.7

(6.4)

161.9

(0.2)

27.1

(0.0)

25.0

(0.8)

26.1 (1.1) 23.7

(0.1)

22.7

(0.1)

23.2 (0.5)
LSJ LAT 102 191 293 47.5

(0.5)

53.1

(1.1)

50.3

(2.8)

71.2

(0.9)

65.3

(3.0)

68.3

(3.0)

162.1

(1.0)

161.7

(6.4)

161.9

(0.2)

27.1

(0.0)

25.0

(0.8)

26.1 (1.1) 34.7

(5.1)

34.9

(9.1)

34.8 (0.1)
Thoracolumbar PA 47 80 127 29.8

(6.8)

46.6

(7.5)

38.2

(8.4)

54.9

(0.9)

59.5

(2.8)

57.2

(2.3)

159.3

(5.1)

152.3

(7.3)

155.8

(3.5)

21.9

(2.4)

26.0

(3.7)

24.0 (2.1) 21.0 (0.5) 22.4

(1.1)

21.7 (0.7)
Thorax PA 56 43 99 34.5

(8.5)

35.9

(2.1)

35.2

(0.7)

53.8

(0.1)

58.0

(7.2)

55.9

(2.1)

161.9

(6.6)

155.7

(8.7)

158.8

(3.1)

20.8

(2.7)

24.5

(5.7)

22.7 (1.9) 20.7

(0.7)

21.8

(2.0)

21.3 (0.6)
LS AP 79 115 194 45.6

(3.5)

38.8

(0.3)

42.2

(3.4)

60.9

(0.7)

63.2

(4.9)

62.1

(1.2)

167.1

(4.1)

158.7

(0.6)

162.9

(4.2)

21.9

(0.9)

25.1

(1.8)

23.5 (1.6) 21.6

(0.2)

22.5

(0.8)

22.1 (0.5)
LS LAT 79 115 194 45.6

(3.5)

38.8

(0.3)

42.2

(3.4)

60.9

(0.7)

63.2

(4.9)

62.1

(1.2)

167.1

(4.1)

158.7

(0.6)

162.9

(4.2)

21.9

(0.9)

25.1

(1.8)

23.5 (1.6) 33.4

(2.6)

34.4

(3.8)

33.9 (0.5)
Shoulder AP 63 52 115 24.1

(2.9)

29.1

(6.1)

26.6

(2.5)

54.4

(0.8)

55.9

(4.4)

55.2

(0.8)

156.0

(6.3)

152.5

(4.5)

154.3

(1.8)

22.5

(2.1)

23.8

(2.9)

23.2 (0.7) 21.1

(0.6)

21.5

(1.1)

21.3

(0.2)

Ankle AP 262 333 595 45.5

(7.5)

33.6

(7.6)

39.6

(6.0)

59.2

(3.0)

58.9

(9.4)

59.1

(0.2)

158.7

(3.7)

155.1

(7.1)

156.9

(1.8)

23.6

(2.3)

25.0

(6.2)

24.3 (0.7) 21.8

(0.8)

22.0

(2.3)

21.9 (0.1)
ALL/Range 1483 2104 3587 21.2- 47.5 29.1- 53.1 26.6-50.3 51.4-71.2 50.6-65.3 54.4- 68.3 154.- 167.1 146.0- 161.7 152.3- 162.9 20.8- 25.3 20.2 – 27.0 21.0- 26.1 8.1- 34.7 6.7- 34.9 7.4- 34.8

Note: De is the radiation thickness of penetration through patient skin during clinical examinations.

Table 3: Average radiographic parameters setting during examination by sex, averaged over sexes and radiograph, presented in mean and SEM (in bracket)        

Radiograph kVp mAs FSD (cm) FFD (cm) Avg. A(FFD) (cm2) (Range)
M F B M F B M F B M F B
Chest( PA) 8.7

(6.4)

80.9

(6.4)

79.8

(1.1)

38.7

(13.4)

37.0

(11)

37.9

(09)

27.6

(0.4)

128.5

(0.3)

128.1

(0.5)

150.0

(0.0)

150.0

(0.0)

150.0

(00)

32.0²

(5.2²-34.2²)

Chest( AP) 5.3

(0.0)

76.2

(0.0)

75.8

(0.5)

18.3

(0.0)

14.8

(0.6)

16.6

(1.8)

78.6 (0.0) 79.8

(0.0)

79.2

(0.6)

100.0

(0.0)

100.0

(0.0)

100.0

(0.0)

30.0²

(5.2²- 34.2²)

Abdo. (AP) 77.4

(3.6)

76.9

(1.7)

77.2

(0.3)

49.6

(13.4)

52.7

(15.3)

51.2

(1.6)

77.5 (0.5) 77.3

(1.2)

77.4

(0.1)

100.0

(0.0)

100.0

(0.0)

100.0

(0.0)

23.2²

(17.1²-38.5²)

Pelvis (AP) 79.3

(5.7)

82.1

(6.0)

80.7

(1.4)

47.9

(10.4)

51.7 (9.4) 49.8 (1.9) 71.3 (3.5) 78.5

(2.4)

70.2

(8.4)

90.0

(0.0)

100.0 (0.0) 950.0

(5.0)

28.3²

(25.0²- 32.0²)

Pelvis (LAT) 85.2

(4.9)

87.1

(5.0)

86.2

(1.0)

42.0

(0.0)

43.6 (1.6) 42.8 (0.8) 65.5 (0.3) 68.3

(2.1)

66.9

(1.4)

100.0

(0.0)

100.0 (0.0) 100.0

(0.0)

23.7²

(12.5²- 17.1²)

Skull (AP/PA) 70.2

(1.9)

70.6

(5.5)

70.4

(0.2)

40.1

(1.9)

39.2 (3.1) 39.7 (0.5) 71.2 (2.1) 72.1

(5.1)

71.7

(0.5)

90.0

(0.0)

90.0

(0.0)

90.0

(0.0)

21.5²

(17.3²- 25.6²)

Skull (LAT) 67.6

(7.6)

75.1

(0.0)

72.9

(3.8)

47.7

(2.7)

50.3 (0.0) 49.0 (1.3) 73.9 (6.0) 78.8

(0.0)

76.4

(0.6)

90.0

(1.8)

90.0

(5.0)

90.0

(5.0)

21.5²

(17.3²- 25.6²)

Hand (AP) 62.6

(2.5)

64.1

(1.0)

63.4

(0.8)

31.9

(5.9)

33.0 (4.9) 32.5 (0.6) 80.7 (12.9) 80.6

(13.8)

80.7

(0.1)

95.0

(5.0)

95.0

(5.0)

95.0

(0.0)

9.3

²(3.5²- 13.2²)

Knee-Joint (LAT) 63.0 (7.4) 61.7

(10.4)

62.4

(0.7)

21.2

(10.9)

23.2

(11.2)

22.2 (1.0) 75.3 (14.7) 78.7

(14.9)

77.0

(1.7)

90.0

(5.0)

90.0

(5.0)

90.0

(0.0)

11.5²

(8.9²- 15.9²)

LSJ (AP) 75.0

(15.0)

78.1

(14.1)

76.6 (1.6) 51.8 (16.4) 50.9 (20.3) 51.4 (0.5) 76.4 (0.1) 77.4

(0.1)

76.9

(0.5)

100.0 (0.0) 100.0 (0.0) 100.0 (0.0) 35.0²

(32.0²- 42.0²)

LSJ (LAT) 80.5 (3.5) 81.8 (4.8) 81.2 (0.4) 56.9 (21.5) 57.8 (27.2) 57.4

(0.5)

65.3 (0.8) 65.1

(1.3)

65.2

(0.1)

100.0 (0.0) 100.0 (0.0) 100.0 (0.0) 30.5²

(27.0²- 32²)

Thoraco- Lumbar (PA) 72.3 (0.2) 72.5 (3.8) 72.4 (0.1) 20.4

(0.9)

24.2 (2.9) 22.3 (5.9) 76.8 (1.8) 78.3

(1.7)

77.6

(0.8)

100.0 (0.0) 100.0 (0.0) 100.0 (0.0) 19.3²

(10.0²- 14.5²)

Thorax (PA) 71.6 (0.8) 74.3 (1.2) 73.0 (1.4) 32.0

(0.0)

27.2 (3.5) 29.6 (2.4) 104.2 (25.9) 103.2

(27.0)

103.7

(0.5)

125.0

(25.0)

125.0 (25.0) 125.0 (0.0) 25.5²

(20.0²- 30.5²)

LS (AP) 85.4 (5.4) 85.6 (5.6) 85.5 (0.1) 61.8 (16.5) 61.8 (13.6) 61.8 (0.0) 78.5 (0.2) 77.5

(0.8)

78.0

(0.5)

100.0 (0.0) 100.0 (0.0) 100.0 (0.0) 32.7²

(28.0²- 42.5²)

LS (LAT) 90.1 (4.1) 90.5 (5.5) 90.3 (0.2) 65.2 (14.8) 68.4 (14.7) 66.8 (1.6) 66.6 (0.3) 65.6

(1.7)

66.1

(0.5)

100.0 (0.0) 100.0 (0.0) 100.0 (0.0) 30.5²

(27.0²- 34.0²)

Shoulder

(AP)

59.9 (0.3) 60.8 (0.1) 60.4 (0.5) 19.9

(0.4)

21.8 (3.3) 20.9 (1.4) 73.7 (1.5) 72.2

(2.1)

73.1

(0.9)

90.0

(0.0)

90.0

(0.0)

90.0

(0.0)

15.5²

(10.5ˉ²-17.3²)

Ankle (AP) 60.0 (3.4) 61.9 (4.3) 61.0 (1.0) 19.2

(3.1)

21.0 (4.1) 20.1 (0.4) 78.1 (1.6) 77.9

(1.2)

78.0

(0.1)

90.0

(0.0)

90.0

(0.0)

90.0

(0.0)

14.2²

(10.3²-18.0²)

 Table 4: Estimated parameters (Dose descriptors and Effective dose) by sex, averaged over sexes and radiographs

      Radiograph ESD (mGy) ESD DAP (mGy cm²) DAP ED (mSv) ED/DAP (mSv/Gycm2)
  M F B (SEM) RF M F B (SEM) RF M F B (SEM) M F B
Chest (PA) 0.740 0.736 0.738 (0.002) 13 1718.7 1634.2 1676.5 (42.3) 21 0.117 0.113 0.125 (0.008) 0.068 0.069 0.069 (0.001)
Chest (AP) 0.885 0.711 0.798 (0.087) 17 750.0 606.6 678.3 (71.7) 39 0.237 0.205 0.221 (0.016) 0.316 0.338 0.327 (0.011)
Abdo. (AP) 2.483 2.618 2.551 (0.068) 21 1157.8 1230.2 1194.0 (36.2) 21 0.354 0.374 0.364 (0.010) 0.306 0.304 0.305 (0.001)
Pelvis (AP) 2.974 2.839 2.907 (0.068) 10 1663.8 1795.8 1729.8 (66.0) 29 0.352 0.356 0.354 (0.002) 0.212 0.198 0.205 (0.007)
Pelvis (LAT) 3.567 3.559 3.563 (0.004) 11 1043.4 1083.2 1063.3 (19.9) 15 0.018 0.022 0.020 (0.002) 0.017 0.020 0.019 (0.002)
Skull (AP/PA) 1.996 1.924 1.960 (0.036) 22 819.9 801.5 810.7 (9.2) 18 0.011 0.013 0.012 (0.001) 0.013 0.016 0.015 (0.002)
Skull (LAT) 2.004 2.293 2.149 (0.145) 19 975.2 1028.4 1001.8 (26.6) 11 0.006 0.009 0.008 (0.002) 0.006 0.009 0.008 (0.002)
Hand (AP) 0.964 1.048 1.006 (0.042) 9 122.0 126.2 124.1 (2.1) 10 0.013 0.014 0.014 (0.001) 0.107 0.111 0.109 (0.002)
Knee Joint (LAT) 0.831 0.875 0.853 (0.022) 5 124.1 135.7 129.9 (5.8) 13 0.001 0.001 0.001 (0.000) 0.008 0.007 0.008 (0.001)
LSJ (AP) 2.506 2.601 2.554 (0.048) 18 2752.0 2704.2 2728.1 (23.9) 22 0.471 0.555 0.513 (0.042) 0.171 0.205 0.188 (0.017)
LSJ (LAT) 4.341 4.581 4.461 (0.120) 21 2295.6 2331.9 2313.8 (18.2) 16 0.131 0.177 0.154 (0.023) 0.057 0.076 0.067 (0.010)
Thoracolumbar (PA) 0.908 1.042 0.975 (0.067) 16 336.2 398.7 367.4 (31.3) 14 0.007 0.006 0.007 (0.001) 0.021 0.015 0.018 (0.003)
Thorax (PA) 0.758 0.708 0.733 (0.025) 6 920.3 782.3 851.3 (6.9) 19 0.045 0.041 0.043 (0.003) 0.049 0.052 0.051 (0.002)
LS (AP) 3.672 3.785 3.729 (0.057) 23 2865.9 2865.9 2865.9 (0.0) 26 0.828 0.815 0.822 (0.007) 0.289 0.284 0.287 (0.003)
LS (LAT) 5.990 6.535 6.263 (0.273) 23 2630.5 2759.6 2695.1 (64.6) 21 0.367 0.378 0.373 (0.006) 0.140 0.137 0.139 (0.002)
Shoulder (AP) 0.660 0.776 0.718 (0.058) 5 211.5 231.7 221.6 (10.1) 25 0.006 0.007 0.007 (0.001) 0.028 0.030 0.029 (0.001)
Ankle (AP) 0.569 0.666 0.618 (0.049) 8 171.2 187.3 179.3 (8.1) 16 0.001 0.001 0.001 (0.000) 0.006 0.005 0.006 (0.001)

Table 5: Average Tissue dose equivalent and the probability of fatal risk of cancer incidence by sex, averaged over sexes and examination

Radiograph Tissue Dose Equivalent (HT) (mSv) Probability of fatal risk of cancer x 10-5 Lifetime cancer risk equivalent  [Averaged over sexes (Both)] Risk category
M F B(SEM) M F B
Chest (PA) 0.784 0.780 0.783(0.002) 4.314 4.291 4.303 1 : 23,240 Very Low
Chest (AP) 0.938 0.754 0.846(0.092) 5.160 4.145 4.653 1 : 21,492 Very Low
Abdo. (AP) 2.632 2.775 2.704(0.072) 14.476 15.263 14.870 1 :   6,725 Low
Pelvis (AP) 3.152 3.009 3.081(0.072) 17.338 16.551 16.945 1 :   5,902 Low
Pelvis (LAT) 3.781 3.773 3.777(0.043) 20.796 20.749 20.772 1 :   4,814 Low
Skull (AP/PA) 2.074 2.000 2.037(0.037) 11.409 11.0011 11.205 1 :   8,925 Low
Skull (LAT) 2.124 2.431 2.277(0.153) 11.683 13.368 12.526 1 :   7,983 Low
Hand (AP) 1.022 1.111 1.066(0.045) 5.620 6.110 5.865 1 : 17,050 Very Low
Knee Joint(LAT) 0.881 0.928 0.904(0.023) 4.845 5.101 4.973 1 : 20,109 Very Low
LSJ (AP) 2.656 2.757 2.707(0.050) 14.610 15.164 14.887 1 :   6,717 Low
LSJ (LAT) 4.602 4.856 4.729(0.127) 25.311 26.708 26.010 1 :   3,845 Low
Thoracolumbar(PA) 0.963 1.105 1.034(0.071) 5.294 6.075 5.684 1 : 17,593 Very Low
Thorax (PA) 0.804 0.751 0.777(0.027) 4.419 4.128 4.274 1 : 23,397 Very Low
LS (AP) 3.892 4.012 3.952(0.060) 21.408 22.067 21.737 1 :   4,601 Low
LS (LAT) 6.349 6.927 6.383(0.290) 34.920 38.099 36.510 1 :   2,739 Low         (H)
Shoulder (AP) 0.700 0.823 0.761(0.062) 3.848 4.525 4.186 1 : 23,889 Very Low
Ankle (AP) 0.603 0.823 0.655(0.052) 3.317 3.883 3.600 1 : 27,778 Very Low (L)

 Note: Risk coeff.(fr) = 5.5 x 10-5 mSv-1 used in this study for probability of fatal risk of cancer incidence calculation (Proposed nominal life-time radiation risk ranged as (2-10 x 10-2 Sv-1) [17][18]

Table 6: Comparison of the average estimated parameters for reference person studied with some published values

Radiographs This Study Ref.[33] Ref.[34] Ref. [32]
  ESD (mGy) DAP (Gycm2) ED (mSv) ESD (mGy) DAP (Gycm2) ESD (mGy) DAP (Gycm2) ED (mSv)
Chest (PA) 0.738 1.676 0.125 0.16 0.12 3.01 3.06 0.42
Chest (AP) 0.798 0.678 0.221 0.16
Abdo. (AP) 2.551 1.194 0.364 4.70 3.00 5.67 17.16 1.82
Pelvis (AP) 2.907 1.730 0.354 4.20 3.00 2.84 3.28 0.52
Pelvis (LAT) 2.727 1.063 0.020
Skull (AP/PA) 1.960 0.811 0.021 2.30 3.93 4.53 0.14
Skull (LAT) 2.149 1.002 0.008 1.35
Hand (AP) 1.006 0.124 0.014 1.44 0.92
Knee Joint (LAT) 0.853 0.130 0.001 0.29 1.73 1.53
LSJ (AP) 2.554 2.728 0.513 28.10 3.00
LSJ (LAT) 4.461 2.314 0.154
Thoracolumbar (PA) 0.975 0.367 0.007
Thoraxis (PA) 0.733 0.851 0.043
LS (AP) 3.729 2.866 0.822 5.90 1.60 3.79 2.72 0.66
LS (LAT) 6.263 2.695 0.373 14.00 3.00
Shoulder (AP) 0.718 0.222 0.007 0.19
Ankle (AP) 0.618 0.179 0.001

DISCUSSION

This study examined 3,587 adult patients undergoing most frequent radiographs involving 17 different procedures from 17 health care centres. These healthcare centres were divided into groups which comprises of public (Group A) and private (Group B). Table 1(a&b) revealed specific features of x-ray units included in the study. The beam output (µGy(mAs)-1 ) of the investigated centre’s x-ray machine range between 25.05 and 74.28 µGy (mAs)-1 and their filtrations range between 1.5 to 2.5 mmAl. Among the x-ray machines studied, twelve satisfied the minimum filtration requirement of 2.5 mmAl, for a good practice [22], for machines operating at a peak voltage of 70 kV [23]. However, use of filtration below the minimum legal requirement of 2.5 mmAl for peak tube potential values greater than 70 kV [23], may have contributed as a factors to relatively higher patient doses. Reference to some studies as, [24] filtration recorded were [(UCH:2.7, OAUTHC:1.7, TDC:2.7) mmAl, NHA:(1.0 mmAl+0.1mmCu)] [25], (H1: 3.0, H2: 1.0, H3: 2.0 mmAl) indicated that the case is not just limited to this study area alone but the country centres at large used relatively low filtration compared to the filtrations reported in Sudan [8], ranged between 2.5 and 5.0 mmAl, and the age of the x-ray machines assessed based on their year of installation, range between 18 and 22yrs.

Total of 3,587 dose measurements on most frequent radiographs were recorded during study. The gender ratio were 41.3% and 58.7%  for males and females respectively as shown in Table 2, and this may likely be reflecting the demography of the study area (Ondo State), Nigeria. The average age, weight and BMI of patients studied ranged (21.2 – 53.1) yrs., (50.6 – 71.2) kg and (20.2 – 27.0) kgm-2. So, the mean weight recorded from this study is far different from those recorded in the IAEA study [12] on patient undergoing radiographs in European and Asian countries. In IAEA study, average weight (70 ± 10) kg and (65±10) kg were considered appropriate for European and Asian countries respectively while the average for the only African country included in the study was not specified. So, it is relevant to compare adequately, the estimated parameters in the study with the published values based on the average weight recorded in Nigeria. The depths of beam penetration recorded ranged (6.7-3.9) cm, reflects the thicknesses of the area exposed and average mass-energy absorption coefficient’s estimating distant for each examination, and so, the real-time radiological practice for the state, deduced. It is equally evident from the average BMI range of the patient recorded, that the average size of the patient exposed to different radiological examinations are closely related (range factor of 1.35).

The summary of the average radiographic exposure data used during radiological examination are reported (Table 3). The average technical factors selection (kVp, mAs, FSD, FFD and the field size area) as determined for study, ranged as (59.9 – 90.5) kVp, (19.2 – 68.4) mAs, (90 -150 cm) FFD and (9.32 – 35.02 cm2) for the beam size. So, averagely low kVp and mAs were recorded from Shoulder AP and Ankle AP and high for LS LAT. Therefore, study revealed inconsistencies in the use of FFD compared to EC quality criteria, recommendations [26].The EC criteria recommends an average FFD of 115 cm and ranged (100-150) cm, but the hospitals studied used FFD below the recommended averaged except in Chest PA, where the FFD used cut across. As entrance surface dose (ESD) is considered an inverse proportionality of the square of FFD, if EC criteria is considered the yardstick, then, for same kV and mAs, dose reaching the patient skin surface is expected to be high. But the observable trends across centres studied tend to using lower FFDs, and in part, this may illuminate higher ESDs. Though, possibility of some centres using low FFDs, presenting considerably low mean ESD cannot be ruled out. It is therefore, expedient to know that changing FFD at time may be seen as good effort, but will still not unravel all the discrepancies noted in this study. So, it is ideal that compliance to policies on quality control and assurance monitoring programs, be adhere and enforce in the hospitals to protect patient against irrational exposure from repeat examinations. However, if images are considered created and ranked in term of signal-noise ratio (SNR) [27], from low to high quality, with exposure factors selection varying as (75 -105) kVp, (20 -50) mAs and FFD (110cm) for standard adult (diameter ‘D’≥20 cm) during examination, together with some identified type of films [28][29], then averagely high SNR may be generated from almost all the centres studied. Hence, averagely high image quality may be expected.

The averaged over centres estimated parameters (ESD, DAP and ED) across radiographs studied by sex and averaged over sexes are reported (Table 4). Generally, along the line of data analysis by centre, considerable variations noted between centres for same type of examinations, even from same room. When values were made averaged over centres for each radiograph, the ESD values across gender, reflects low and high standard error of the mean (SEM) being 0.002 and 0.273 (highlights of spread across radiograph) from Chest PA and LS LAT radiographs, and the range factor across all centre studied for each radiograph, show moderately high spread with low and high values recorded approximately as 5 (Knee Joint LAT and Shoulder AP) and 23 (LS generally). These variations were still comparable to some studies. Generally, across radiographs, low and high, averaged over sexes ESD recorded from Ankle PA (0.618 mGy) and LS LAT (6.263 mGy) with range factor of 10.1 across radiographs. These shows geometry agreement with IAEA study [12], but the difference in values may be ascribed to technical factors selections, patient sizes, x-ray machine beam output, and dosimetric method adopted. Also, low and high averaged over sexes DAP were recorded from Hand AP (0.124 Gycm2) and LS AP (2.866 Gycm2) while  for  the ED,  Knee joint (LAT) and Anxle AP recorded same low value of 0.001 mSv and LS AP recorded high value of 0.822  mSv. The low and high ED/DAP, made averaged over sexes was 0.006 and 0.327 mSv/Gycm2 from Ankle AP and Chest AP respectively.

The average probability of fatal risk of cancer incidence by sex and averaged over sexes and examination presented in Table 5. So, low and high probability of fatal risk of cancer incidence, averaged over sexes recorded as (3.600 and 36.510) x 10-5 from Ankle AP and LS LAT respectively while the low and high Lifetime cancer risk equivalent ratio was 1: 27,778 and 1: 2,739 from Ankle AP and LS LAT respectively. Therefore, from study, it is evident that the Lifetime risk of cancer equivalent recorded fall within low (1:1,000) and very low (1:10,000) category of risk, as indicated in four broad risk band [30]. This notwithstanding do not justified the best radiological practice, but the need for adequate monitoring of dose to patients for further optimization of the practice.

Comparison of the estimated parameters from this study with some published values, established in Table 6. The observable trend from this study do not necessarily mean the best practice but calls for the establishment of legal framework on dose optimization and overview of same in the entire healthcare within the state, adopting direct dosimetry method for diagnosis purposes. Nonetheless, dose recorded from this study when compared with other published values, show geometric agreement, but the noticeable difference may be ascribed to technical factors selection, patient sizes, radiological practice, x-ray beam output generation and the choice of dosimetry method adopted. However, it has been established that where quality dose database are used in dose to patient monitoring, the difference in dose values between direct and indirect dosimetry will range as (±20% and 25%) [31]. Thus, the need for the variations recorded between this study and others, published. So, the range factors (RF) of the estimated parameters from this study were 10.10, 15.94 and 820 for ESD, DAP and ED respectively which signified high spread in values, especially in ED compared to [32] and [34] with RF of 2.73, 18.65 and 13.00 and [33] with RF as 175.63 and 25.00 for ESD and DAP respectively.  Therefore, observable trend justified expected variation in dose to patient from radiographic rooms or across centres [9], which are traceable to patient sizes, technical factors selection and adopted dosimetric method. So, to guarantee adequate radiation protection of patients, dose to patient through diagnostic exposure in the state, be optimized periodically to promote good radiologic practice. Hence, the need for adequate monitoring in terms of the personnel, equipment, and techniques of exposure, and dosimetric exercise, embraced.

CONCLUSION

The results presented in this study, provide current state of radiological practice in the study area. It displayed variation in technique, exposure factors and radiation doses for same and the different type of radiographs studied, which strongly support the need for further optimization.  Almost seventy percent of the centres studied recorded low ESD and DAP values below UK national reference values with high spread in dose to patients. Hence, the urgent need for appropriate and realistic QA program to be implemented, which is currently absent in almost all the hospitals surveyed.  The variation in the data obtained, equally validates the importance for awareness creation among the radiographic staff and the policies maker on regular quality control testing of equipment and standardization of protocols.  Thus, the need for intervention and appropriate corrective actions to improve and standardized practice, enhancing the quality of the radiographs, and avoid illogical risks of increased radiation dose to patients and staff, advised. Therefore, bringing the standard of the radiological practice in the state in line with national and European guidelines on quality criteria for diagnostic radiological images becomes highly imperative, in order to protect patients and staff against irrational exposure.

ACKNOWLEDGMENTS

Authors would like to acknowledge with gratitude the sponsor of this work through institutional based research (IBR) funding, and equally appreciate the services and support provided by all the staff at the radiology department of each studied centres, during the period of data collection.

REFERENCES

  1. Williams and wilkins (2002). Radiation dose associated with common computed tomography. Egyptian J. of radiology and Nuclear Medicine (https://doi.org/10.1016/j.ejrnm)
  2. Howell, JD (2016). Early clinical use of the x-ray. J. of America Clinical and Climatological Association. 127 (2016): 341-49.
  3. United Nation Scientific Committee on the Effect of Atomic Radiation (UNSCEAR) (2000). Source and effect of ionizing radiation (Report to the General Assembly with scientific annexes), 2000 UN; New York.
  4. Tsapakpi V, Tsalafoutas A, Chinofoti I (2007). Radiation dose to patient undergoing some common radiographic examinations in North Africa. NRPB report (HMSO), New York, USA.
  5. Parry RA, Glaze SA, Archer BR (1999). Typical patient radiation doses in diagnostic radiology. AAPM/RSNA physics tutorial for residents (Imaging and the therapeutic technology)
  6. National Radiation Protection Board (NRPB) (1990). General and Miscellaneous (F6000); radiation protection on dental (C5500), NRPB annual report 1990-91; Chilton, UK
  7. National Occupational Health Safety Commission (NOHSC) (1995). Guidance note on the interpretation of exposure standards for atmospheric contaminations in the occupational environment. NOHSC-3008, 3rd edition (Update for amendment)
  8. Suliman II, Abbas N, Habbani FI (2007). Entrance surface doses to patient undergoing selected diagnostic x-ray examinations in Sudan. Radiat prot dosimetry 123(2); 209-14
  9. Aweda MA (2005). Evaluation of patient skin effective dose due to diagnostic procedures with x- rays in Lagos state, Nigeria. Nigerian J. of Health and Biomedical Sciences. 4(1); 46-52.
  10. Schandford C and Tetteh GK (1998). Analysis of dose and dose distribution for patient undergoing selected x-ray diagnostic procedures in Radiat prot. Dosimetry, 76(4); 249-56.
  11. Ondo State bureau of statistics (2006) (http://www.ondostatistics.org/ondoprofile.php)
  12. International atomic energy agency (IAEA) (2004). Optimization of the radiological protection of patients undergoing radiography, fluoros copy and computed tomography. Report of a coordinated research project in Africa, Asia and Eastern Europe. IAEA- TECDOC-1423, Vienna, Austira, IAEA; 2004.
  13. Tung CJ, Tsai HY (1999). Evaluation of gonad and fetal doses for diagnostic radiology. Proceedings of the national sciences council, Republic of China, Patr.B, life sciences:23(3);107-13
  14. Faulkner K, Broadhead DA, Harrison RM (1999). Patients dose measurement methods. J.Applied radiation and Isotopes 50(4):113-23.
  15. Jones DG and Wall BF (1985). Organ doses from medical x-ray examination. Radiat. Prot. Dosimetry. 57(1): 643-67
  16. Meade AG, Osibote AO, Faulkner K (2003). Assessing common methods based on organizational research. Paper presented at 22nd annual meeting of the society of industrial organizational psychology, New York: 3-15
  17. ICRP. The 2007 Recommendations of the International Commission on Radiological protection. ICRP publication 103. Ann. ICRP 37, 2007
  18. Bushra, E., Sulieman, A. and Osman, H (2010). Radiation measurement and risks estimation for paediatric patients’ during routine diagnostic examinations. In: Tenth Radiation Physics and Protection Conference, Nasr City, Cairo (http://www.researchgate.net/publication/267816964)
  19. Balter S, Fletcher DW, Kuan HM, Miller D, Richter D, Seissl H, Shope, Jr. TB (2002). Techniques to estimate radiation dose to skin during fluoroscopically guided procedures. Skin Dose Measurements, AAPM, 2002 (July): 1-10
  20. American Association of Physicist in Medicine (AAPM) (1988). Protocol for the radiation safety survey of diagnostic radiological equipment (AAPM no:25); J. of AAPM; 1988: 214-301
  21. Institute of physics and engineering in medicine (IPEM) (2005). Recommended standards for the routine performance testing of diagnostic x-ray imaging system. IPEM Report 91, York, UK.
  22. Johnston DA and Brennan PC (2000). Reference dose levels for patient undergoing common diagnostic X‐ray examination in Irish hospitals. Br J Radiology.73 (868): 396–402.
  23. Wade JP, Goldstone KE, Dendy PP (1995). Measurement and dose reduction in East Anglia (UK). Radiat Prot Dosimetry: 57(1- 4):445-48
  24. Akinlade BI, Farai IP, Okunade AA (2012). Survey of dose area product received by patients undergoing common radiological examinations in four centers in Nigeria.J.Applied clinical medical physics (spring).13(4):188-96.
  25. Olowookere CJ, Obed RI, Babalola IA, Bello TO (2011). Patient dosimetry during Chest, Abdomen, Skull and Neck radiography in Southwest, Nigeria. Radiography, 17(3); 245-49
  26. European Commission. European guidelines on quality criteria for diagnostic radiographic images. EUR.16260 EN. Luxembourg: OOPEC, 1996.
  27. Bandura A (1997). Self-efficacy: The exercise of control. New York; Worth publisher
  28. Carver B and Carver B (2012). Medical imaging techniques, reflection and evaluation (2nd Edition), Edinburgh; Churchill Livingstone
  29. Whitley AS, Sloane C, Hoadley G, Moore AD, and Alsop CW (2005). Clark’s positioning in radiography (12 Edition). London: Hodder Arnold
  30. Wall BF, Haylock R, Jansen JTM, Hillier MC, Hart D and Shrimpton PC (2011). Radiation risks from medical x-ray examinations as a function of the age and sex of the patient. Report HPA-CRCE-028. Health Protection Agency: Chilton.
  31. Ofori EK, Antwi WK, Scutt DN, Ward M (2012). Optimization of patient radiation protection in pelvis x-ray examination in Ghana. J. Applied clinical medical physics (spring);13(4):1-12
  32. Jibiri NN and Olowookere, CJ (2016). Evaluation of dose-area product of common radiographic examinations toward establishing a preliminary diagnostic reference level (PDRLs) in Southwestern, Nigeria. J. Applied Clinical Med. Phys.;17(6):392-404 (doi:10.1120/jacmp.v17i6.6011)
  33. Hart D, Hillier MC, Wall BF (2002). Dose to patient from medical x-ray examinations in UK-2002 reviews. NRPB –W14. Chilton, UK: NRPB; 2002
  34. Jibiri NN and Olowookere CJ (2016). Patient dose audit of the most frequent radiographic examinations and the proposed local diagnostic reference levels in southwestern, Nigeria: Imperative for dose optimization (doi:org10.1016/jjrras.2016.01.003) (www.sciencedirect.com)

Article Statistics

Track views and downloads to measure the impact and reach of your article.

0

PDF Downloads

598 views

Metrics

PlumX

Altmetrics

GET OUR MONTHLY NEWSLETTER